Hypercure
Overview
  • The spectral transmission curve in the brochure represents output with a liquid light guide which limits transmission to above 270nm.

  • The precision coated elliptical mirror in the h200 lamp housing will fully reflect 250 - 550nm. The coating on the elliptical mirror eliminates infrared heat transmission, furthermore an optical filter-holder is built-in to the hypercure to accommodate optional band-pass filters and cold filters.

  • The 200 watt high pressure Hg-Xe lamp outputs above 250nm with spectral peaks at 310, 365, 410, and 435nm.

  • For high-output intensity in the 254 - 310nm range a quartz or high grade fused silica light guide is recommended (for square area exposure a quartz integrator is recommended).

  • The intensity output of the lamp over its lifetime remains above 80% with an estimated lifetime of 1000 hrs.

  • Fiber optic light guides may be bundled with specific or random fiber patterns for particular applications.

  • Intensity output significantly decreases as the distance between the sample and light output point increases.

  • Intensity output with the uniform area exposure adaptor will increase with a reduction in the exposure area.

  • Precise distance must be maintained for intensity uniformity:
2" x 2"
3" x 3"
4" x 4"
90mW/cm2
60mW/cm2
40mW/cm2
+/- 4%
+/- 4%
+/- 4%
  • 300 Watt hypercure systems are available on special order.
Back