| MX 301-AZ | ||||||||||||||
| Automatic Thickness Measurement System | ||||||||||||||
|
|
![]() |
|||||||||||||
| This gauge is used to measure the center thickness of 75mm, 100mm, 150mm and 200mm silicon wafers. | ||||||||||||||
| Specifications | ||||||||||||||
| Resolution
Noise Precision (Repeatability) Overall Accuracy |
10 nm |
|||||||||||||
| *within a temperature range of ±5 K and a thickness range of ± 100 µm no recalibration or of the system is required ! | ||||||||||||||
| Under
control of a computer or an automation system connected to the MX 301-AZ
by means of a serial interface, a robot hand takes single wafers out of
any slot of a standard wafer carrier, feeds them into the measuring gauge,
and puts them back into the same slot after the measurement. The system
can be set up to measure all wafers available in a carrier, or to pick a
freely selectable subset of wafers out of it. In this case, the host must
transmit the type of wafer carrier in use, and before every measurement,
the number of the carrier slot to be addressed.
The system offers three modes of operation:
SPC (statistical process
control) software is available for the evaluation of the sampled data.
|
||||||||||||||