| MX 102-6, MX 102-8 | |||||||||||||||
| High Resolution Wafer Thickness and Thickness Variation Gauge | |||||||||||||||
| System Description | |||||||||||||||
|
The
MX 102-6 and MX 102-8 are instruments for the measurement of center thickness
and thickness variations of 1OOmm to 150mm and 150mm to 200mm silicon
wafers respectively. Each gauge has a resolution of 10nm and a thickness
range of 300µm to 700µm for the MX 102-6 and 450µm to
850µm for the MX 102-8.
Before each wafer
scan, the gauge automaticaly re-calibrates itself by means of a standard
thickness gauge block. A pair of capacitive sensors samples four radial
profiles of each wafer, each profile measurement points offset by 45°
from the previous profile. A host computer interfaces with the gauges
and records and stores all measurement data. |
|||||||||||||||
| Specifications | |||||||||||||||
| Resolution Spatial Resolution Precision (Repeatability), (wafer still, one measuring point) TTV - Accuracy ± 50 nm Thickness ranges (MX102-6) Thickness ranges (MX102-8) |
10 nm |
||||||||||||||
| Figures 1 and 2 show the mechanical structure of the system. Figures 3 and 4 show the four standard gauge blocks and the exchangeable shims. These shims are coded so the corresponding gauge block is automatically selected. Figures 5 and 6 represent hard copies made during two consecutive scans of the same wafer (demonstration of gauge repetability). The software permits the selection of different scales and offsets of the displayed measurement data. | |||||||||||||||
| Back | |||||||||||||||